dc.creator |
Wassermann Adriana I. |
|
dc.creator |
Gorga Carlos J. P. |
|
dc.creator |
Baumhardt-Neto Ricardo |
|
dc.date |
1998 |
|
dc.date.accessioned |
2013-05-30T10:43:40Z |
|
dc.date.available |
2013-05-30T10:43:40Z |
|
dc.date.issued |
2013-05-30 |
|
dc.identifier |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421998000100016 |
|
dc.identifier |
http://www.doaj.org/doaj?func=openurl&genre=article&issn=01004042&date=1998&volume=21&issue=1&spage=106 |
|
dc.identifier.uri |
http://koha.mediu.edu.my:8181/jspui/handle/123456789/4155 |
|
dc.description |
Infrared spectroscopy laboratories works under permanent care with respect to water contamination, mainly in liquid samples. In this case crystal plates destruction or damage are frequent, increasing the operational expenses. On the other hand, the laboratory which produces such samples must be very careful in drying liquid samples. In this work we develop a simple and inexpensive way to operate in such conditions using polypropylene and HDPE films which were thermally soldered resulting little containers or sample holders. The spectra of sample/sample holder is achieved having the sample holder as background. |
|
dc.publisher |
Sociedade Brasileira de Química |
|
dc.source |
Química Nova |
|
dc.subject |
infrared |
|
dc.subject |
aqueous |
|
dc.subject |
polymer |
|
dc.title |
Porta-amostras polimérico para espectroscopia infravermelha de sistemas líquidos aquosos ou orgânicos |
|