Please use this identifier to cite or link to this item: http://dspace.mediu.edu.my:8181/xmlui/handle/123456789/4190
Full metadata record
DC FieldValueLanguage
dc.creatorMartelli Patrícia B.-
dc.creatorReis Boaventura F.-
dc.creatorZagatto Elias A.G.-
dc.creatorLima José L.F.C.-
dc.creatorLapa Rui A.-
dc.date1998-
dc.date.accessioned2013-05-30T10:47:10Z-
dc.date.available2013-05-30T10:47:10Z-
dc.date.issued2013-05-30-
dc.identifierhttp://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421998000200005-
dc.identifierhttp://www.doaj.org/doaj?func=openurl&genre=article&issn=01004042&date=1998&volume=21&issue=2&spage=133-
dc.identifier.urihttp://koha.mediu.edu.my:8181/jspui/handle/123456789/4190-
dc.descriptionThe construction of a tubular hydrogen ion-selective potentiometric electrode without inner reference solution, based on the tridodecylamine (TDDA) ionophore, and its evaluation in a flow system are described. TDDA was dissolved in 2-nitrophenyl octyl ether, dispersed in a PVC membrane and applied directly to a conducting support which consisted of an epoxy resin and graphite mixture. The electrode was designed with a tubular geometry to effort facilities to be coupled as part of a flow injection network. The main working characteristics such as response time, linear pH range, selectivity and life time were evaluated and compared with those obtained which a conventionally shaped electrode based on the same sensor. The electrode showed a slope of 51-52 mV dec-1 within a linear pH range from 4.0 up to 12.0.-
dc.publisherSociedade Brasileira de Química-
dc.sourceQuímica Nova-
dc.subjecttubular pH electrode-
dc.subjectflow analysis-
dc.subjecttridodecylamine-
dc.titleConstrução e avaliação de um eletrodo tubular sensível ao íon hidrogênio como detector em sistemas de análise em fluxo-